技术文章

碳硫高速分析仪操作[英文]

南京天量智能科技有限公司

2011/3/24 16:47:15

TL-CS型碳硫高速分析仪操作

Carbon & sulfur high speed analysis meter operation (Model:TL-CS)

 

一、打开所有电源和前氧,再开氧气(压力控制在0.02—0.04Mpa )zui后关前氧。

1. Turn on all power supply and frontal oxygen switch,then open the oxygen(the pressure limitation is 0.02—0.04Mpa)shut down the frontal oxygen switch finally.

二、依次加入硅钼粉小半药匙左右,锡粒少量,样件所有普碳钢、低合金铜样件量为1G,分析天平称量到铜坩锅内。

2.   Put half spoon of silicon molybdenum powder in turn,a little tin grain and metal sample(the quantity of carbon steel and low alloy copper sample is 1G, analytical balance quantifyinto copper crucible.

三、按信对零下放,使量气筒液面降到底,调碳调零,显示“0”

3.  Put down switch to the zero scale, lower down liquid level to the bottom of air measuring cylinder, adjust the button“ Carbon to the zero” to the scale“0”.

四、点一下准备,待量气筒、滴定管液面加满调硫调零,显示“0.000”zui后一位千分位可以不为0.

4. Touch the switch preparation, after topping up air measuring cylinder and buret, adjust the button“Sulfur to the zero”to the “0.000”the last number could not be 0.

五、升起升降轴,先开后控再开前氧(控制流量为80-1200.

5.  Rise up the lifting shaft, Turn on, adjust and Turn on frontal oxygen again(adjusted flow area 80-1200)

六、点一下分析后,看集气瓶液面回到原位,后关前氧看D4灯灭时显示数值。(*定标:待D4灯灭即点一下复位,分别调“C”“S”标准使显示标样数值)。

6.Touch the analysis switch, when the liquid level of gas collecting bottle is back to original place shut down the frontal oxygen switch and read the number of D4 light when it is off(Read standard data:touch the reposition, adjust “C”“S” and read standard data.

七、关后控,降下升降轴。

7. turn off back controller,lower down the lifting shaft.

八、关电源与氧气。

8. Shut down the power supply and oxygen.

注:黑体字表示机器按钮或操作手柄。

Note: Overstriking words show the button of machine or operation handle.

 

                                                                                           南京天量分析仪器有限公司

                                                                      2011   03

相关产品