Hipot Test 100 - 5 000 V AC, 0.1 - 100 mA, 500 VA Isc > 200 mA, potential-free Protective Earth Test 0.5 - 25 A AC, 1 - 500 mΩ Insulation resistance 50 - 1 000 V DC, 0.5 - 400 MΩ, current-limited, EN 50191 Test system with HV-AC, ISO, PE Test system with X4 user interface for remote control via PC and for standalone usage: - Test time adjustable from 0,1 s - 99 min, endless - Including manufacturers certificate of calibration Hi Pot test, HV-Module HVA5-100: - Test voltage adjustable from 100 - 5000 V AC, potential free - Trip current adjustable from 0,1 - 100 mA - Frequency adjustable from 45 - 65 Hz - Electronically regulated, incl. ramp function and active discharge - Power of transformer 500 VA, short circuit current > 200 mA - Switching on test voltage at zero crossing of voltage - Monitoring of the test leads for cable breakage - Monitoring of contact to the test object Insulating resistance test, ISO-Module DC10-CL: - Test voltage adjustable from 50 - 1000 V DC - Resistance limits adjustable from 0 - 400 MOhm - Electronically regulated, incl. ramp function and active discharge - Current limited as detailed in EN 50 191 - Monitoring of the test leads for cable breakage - Monitoring of contact to the test object Protective Earth test, PE-Module AC25: - Test current adjustable from 2 - 25 A AC - Resistance limits adjustable from 1 - 500 mOhm - Switchable to test as detailed in EN 60 204 - Test voltage adjustable < 6 V und < 12 V - Electronically regulated Additional output: PEP4-socket - PE-connection for mains terminal of the test object This enables compley independent PE-testing, separated from HV- ISO-testing. |