适用生物 Homo sapiens (Human,人)
淋巴细胞功能关联抗原1α(LFA1α)检测试剂盒检测范围 0.156-10ng/mL 灵敏度 0.054ng/mL
样本类型 Tissue homogenates, cell lysates and other biological fluids.
实验时长 4.5h 实验方法 双抗夹心法 淋巴细胞功能关联抗原1α(LFA1α)检测试剂盒规格 96T
ELISA Kit for Lymphocyte Function Associated Antigen 1 Alpha (LFA1a)
FOR IN VITRO AND RESEARCH USE ONLY, NOT FOR USE IN CLINICAL DIAGNOSTIC PROCEDURES!
Organism species | Homo sapiens (Human) |
Product No. | SEB572Hu |
Sample type | Tissue homogenates, cell lysates and other biological fluids. |
Format | 96-well strip plate |
Assay length | 4.5 hours |
Detection range | 0.156-10ng/mL The standard curve concentrations used for the ELISA’s were 10ng/mL, 5ng/mL, 2.5ng/mL, 1.25ng/mL, 0.625ng/mL, 0.313ng/mL, 0.156ng/mL |
Sensitivity | The minimum detectable dose of this kit is typically less than 0.054ng/mL. |
淋巴细胞功能关联抗原1α(LFA1α)检测试剂盒Specificity
This assay has high sensitivity and excellent specificity for detection of Lymphocyte Function Associated Antigen 1 Alpha (LFA1a).
No significant cross-reactivity or interference between Lymphocyte Function Associated Antigen 1 Alpha (LFA1a) and analogues was observed.
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Lymphocyte Function Associated Antigen 1 Alpha (LFA1a) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Lymphocyte Function Associated Antigen 1 Alpha (LFA1a) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
淋巴细胞功能关联抗原1α(LFA1α)检测试剂盒Stability
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Reagents and materials provided
Reagents | Quantity | Reagents | Quantity |
Pre-coated, ready to use 96-well strip plate | 1 | Plate sealer for 96 wells | 4 |
Standard | 2 | Standard Diluent | 1×20mL |
Detection Reagent A | 1×120μL | Assay Diluent A | 1×12mL |
Detection Reagent B | 1×120μL | Assay Diluent B | 1×12mL |
TMB Substrate | 1×9mL | Stop Solution | 1×6mL |
Wash Buffer (30 × concentrate) | 1×20mL | Instruction manual | 1 |
淋巴细胞功能关联抗原1α(LFA1α)检测试剂盒Assay procedure summary
1. Prepare all reagents, samples and standards;
2. Add 100μL standard or sample to each well. Incubate 2 hours at 37oC;
3. Aspirate and add 100μL prepared Detection Reagent A. Incubate 1 hour at 37oC;
4. Aspirate and wash 3 times;
5. Add 100μL prepared Detection Reagent B. Incubate 30 minutes at 37oC;
6. Aspirate and wash 5 times;
7. Add 90μL Substrate Solution. Incubate 15-25 minutes at 37oC;
8. Add 50μL Stop Solution. Read at 450nm immediay.
淋巴细胞功能关联抗原1α(LFA1α)检测试剂盒Test principle
The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Lymphocyte Function Associated Antigen 1 Alpha (LFA1a). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to Lymphocyte Function Associated Antigen 1 Alpha (LFA1a). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Lymphocyte Function Associated Antigen 1 Alpha (LFA1a), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Lymphocyte Function Associated Antigen 1 Alpha (LFA1a) in the samples is then determined by comparing the O.D. of the samples to the standard curve.