SKC/美国 品牌
代理商厂商性质
上海市所在地
美国SKC个人空气采样泵介绍
通用pcxr8,5至5000毫升/分钟,可编程的个人或地区采样泵可提供高达12小时的运行时间或延长间歇采样颗粒和气体/蒸气。该泵是用于吸附管,理想的过滤盒,大小选择性的采样,样品袋,撞击器。
UL认证本质安全
流量范围:5~5000毫升/分钟。(5~500ml/min 需加装低流量套件)
背压补偿:可达40英吋水柱
运转时间:流量4公升/分钟,20英吋水柱背压,可运转8小时以上
流量精准度:流量校正后±5%
采样前可测试电池状态
可中途暂停采样,保存采样时间
低电量、流量错误自动停机并记忆采样时间
可延迟启动、定时停机、间歇采样
重量:964克,尺寸:4.9×11.9×13公分
内置低流量压力稳定器,能适应多管采样,可设定多管不同流量
RFI/EMI抗无线电波、电磁波干扰之外壳
UL与cUL本质安全认证
ATEX与MSHA认证型号
符合工业卫生和环境应用的方案
美国SKC个人空气采样泵技术参数
Flow Range | 1000 to 5000 ml/min (UL Listed model) |
Weight | 34 oz (964 gm) |
Dimensions | 5.1 x 4.7 x 1.9 in (13 x 11.9 x 4.8 cm) |
Compensation Range | 1000 to 2500 ml/min at 40 inches water back pressure 3000 ml/min at 35 inches water back pressure 4000 ml/min at 20 inches water back pressure 5000 ml/min at 10 inches water back pressure |
Typical Back Pressure | See Table 1. |
Flow Control | Holds constant flow to ± 5% of set-point |
Run Time | NiCad Battery: 8 hrs minimum at 4000 ml/min and 20 inches water back pressure; dependent on media used. See Table 3. NiMH Battery: 12 hrs minimum at 4000 ml/min and 20 inches water back pressure; dependent on media used. See Table 2. Battery Eliminator: Pump will provide extended runs. |
Tubing | Requires 1/4-inch ID tubing |
Timer Display Range | 1 to 9999 minutes (6.8 days). Pump will shut off at 9999 minutes. To reset, restart the pump. |
Flow Indicator | Built-in rotameter with 250-ml division; scale marked at 1, 2, 3, 4, and 5 L/min |
Charging Time | 6 to 8.5 hrs with PowerFlex charger |
Power Supply | 6.0-V plug-in NiMH battery pack, rechargeable, 3.5-Ah capacity or 6.0-V plug-in NiCad battery pack, rechargeable, 2.0-Ah capacity A battery eliminator is available (see Ordering Information); use voids the UL Listing for intrinsic safety. |
Intrinsic Safety | UL Listed for Class I, Division 1 and 2, Groups A, B, C, D; Class II, Division 1 and 2, Groups E, F, G; and Class III, Temperature Code T3C |
Operating Temperature | 32 to 113 F (0 to 45 C) |
Storage Temperature | -4 to 113 F (-20 to 45 C) |
Charging Temperature | 50 to 113 F (10 to 45 C) |
Operating Humidity | 0 to 95% non-condensing |
Flow Fault | If the pump is unable to compensate for longer than 15 seconds due to excessive back pressure, the pump enters flow fault. During flow fault, the pump stops, displays FLOW FAULT, pauses timing functions, and displays elapsed time or pump time. Auto-restart is attempted up to 5 times. |
Low Battery Fault | Pumps stops, displays LO BATT, pauses timing functions, and displays elapsed time or pump time. |
Battery Test | LCD indicator shows battery condition prior to sampling |
Time Display | LCD displays up to 9999 minutes (6.8 days) at which point the display rolls over to 0. Displays include sampler run time in minutes for sampling period elapsed time, pump run time, or total elapsed time including delayed start |
Delay On | Allows user to select minutes to delay start of test up to 9999 minutes (6.8 days) |
Timed Shutdown | Allows user to select minutes of operation before automatic shutdown. Timed shutdown maximum is 9999 minutes (6.8 days) |
Timing Accuracy | ± 0.05% (± 45 sec/day) |
Intermittent Sampling | Programmable to allow user to extend short-term samples over an extended period of time to meet time-weighted average (TWA) requirements with a reduced number of samples. Elapsed time maximum is 9999 minutes (6.8 days), at which time the sample pump shuts down. |
Sampling Pause | Allows user to temporarily halt sampling without loss of timing data. Restart does not require resetting time. |
Multiple-tube Sampling | Built-in constant pressure regulator allows user to take up to four simultaneous samples with rates up to 500 ml/min each (total combined flow 1350 ml/min maximum) using the optional adjustable low flow holder accessory |
RFI/EMI Shielding | Complies with requirements of EN 55022, FCC Part 15 Class B, EN 50082-1 Frequency range of the radiated susceptibility test was 27 to 1000 MHz CE marked |
Additional Specifications | Built-in particulate trap External see-through filter housing Tamper-resistant control cover Tamper-resistant compensation adjustments Built-in accessory mount Stainless steel belt clip External exhaust port All non-plastic parts made of corrosion-resistant materials RFI and EMI-shielded case |
Additional Features | One-year warranty Ultra-quiet operation Pulsation-free flow Impact-resistant case |
配置
表1。
典型背压采样介质(英寸水柱)
信息进行比较,在此回压表泵补偿范围,以确定适当的应用。
过滤器/孔径(μm) | 流量(升/分钟) | ||||
1.0 | 1.5 | 2.0 | 2.5 | 3.0 | |
25毫米MCE/0.8 | 6 | 9 | 12 | 15 | 18 |
25毫米MCE/0.45 | 14 | 22 | 28 | 35 | 40 |
37毫米MCE/0.8 | 2 | 3 | 4 | 5 | 6 |
37毫米PVC/5.0 | 1 | 1 | 2 | 2 | 2.5 |
表2。
泵运行时间的小时,镍氢电池
以下是典型的实现时,使用*充电的镍金属氢化物(镍氢)电池组的运行时间。数据的根据采样介质类型分类。结果均使用一个新的充满电的电池。泵的性能可能会有所不同。
混合纤维素(MCE)过滤器,孔径0.8微米
流量(升/分钟) | 滤镜直径 | |
37毫米 | 25毫米 | |
2.0 | 37 | 33 |
2.5 | 34 | 26 |
3.0 | 31 | 21 |
3.5 | 29 | 18 |
4.0 | 25 | 15 |
4.5 | 20 | 14 |
聚氯乙烯(PVC)5.0微米孔径过滤器,
流量(升/分钟) | 滤镜直径 | |
37毫米 | 25毫米 | |
2.0 | 47 | 41 |
2.5 | 38 | 33 |
3.0 | 35 | 30 |
3.5 | 26 | 27 |
4.0 | 22 | 25 |
4.5 | 21 | 23 |
注意: | 背压在取样过程中由于样品堆积在过滤器上的增加可以降低电池的使用寿命。 |
表3。
泵运行时间的小时,镍镉电池
以下是典型的实现时,使用*充电的镍镉电池(镍镉)电池组的运行时间。数据的根据采样介质类型分类。列出所有的运行时间为小时。
混合纤维素(MCE)过滤器,孔径0.8微米
流量(升/分钟) | 滤镜直径 | |
37毫米 | 25毫米 | |
2.0 | 24.1 | 16.3 |
2.5 | 21.4 | 14.5 |
3.0 | 19.1 | 11.0 |
3.5 | 17.8 | 10.7 |
4.0 | 15.4 | ** |
4.5 | 14.6 | ** |
** | 在测试过程中,过滤器背压超过泵的能力。 |
聚氯乙烯(PVC)5.0微米孔径过滤器,
流量(升/分钟) | 滤镜直径 | |
37毫米 | 25毫米 | |
2.0 | 31.6 | 21.7 |
2.5 | 27.7 | 24.0 |
3.0 | 27.0 | 18.6 |
3.5 | 22.8 | 16.4 |
4.0 | 19.4 | 16.2 |
4.5 | 19.0 | 14.6 |