石英晶体微天平QCM金晶片的校准
时间:2017-04-16 阅读:2701
-
提供商
美国Gamry电化学 -
资料大小
238600 -
资料图片
-
下载次数
61次 -
资料类型
pdf -
浏览次数
2701次
时间:2017-04-16 阅读:2701
提供商
美国Gamry电化学资料大小
238600资料图片
下载次数
61次资料类型
pdf浏览次数
2701次Introduction
This Application Note is the second in a series relating to the eQCM 10M™ and assumes that
you have read and understand the material covered in the Application Note, Basics of a
Quartz Crystal Microbalance.
The purpose of this Application Note is to show the reader a simple experiment where we
reduce Cu2+onto an Au electrode and then oxidize Cu back off. We will then use the result to
calculate a calibration factor for the crystal. Conversely, we will show you how to calculate the
molar mass of the species deposited using a calibration factor.
Each crystal has a theoretical calibration factor, however, under typical experimental
conditions, these calibration factors will vary slightly. We will use Cu2+ to calculate a
calibration factor which could then be used in subsequent experiments on the same crystal.
Cu2+ is reduced in a two electron reaction as shown below:
CuSO4(aq) + 2 e- →Cu(s) + SO42-(aq)
Our goal is to reduce Cu2+ onto an Au electrode, then use the frequency decrease and the
charge passed to calculate a calibration factor.