THDP0100高压差分探头
THDP0100高压差分探头
THDP0100高压差分探头
THDP0100高压差分探头
THDP0100高压差分探头

THDP0100THDP0100高压差分探头

参考价: 面议

具体成交价以合同协议为准
2023-11-10 10:50:08
1030
属性:
产地类别:进口;
>
产品属性
产地类别
进口
关闭
深圳市元锋科技有限公司

深圳市元锋科技有限公司

中级会员12
收藏

组合推荐相似产品

产品简介

THDP0100高压差分探头用于测量两个均非为地的测试点之间的电压差。 泰克的高压差分探头可用于高达 6000 V 的信号。这类探头由于具有共模抑制能力,成为较大部件中进行非地参考、浮动或隔离测量的JIA选择。这些产品由泰克设计、制造和维修。

详细介绍

THDP0100高压差分探头Tektronix offers a broad portfolio of high-voltage probing solutions that enable users to safely and accurately make floating measurements.

Key performance specifications

Key features

Applications

Connectivity

Safe high-voltage probe solutions

The THDP0100 and P5210A have the largest differential dynamic range capability from Tektronix, allowing users to safely measure up to ±6000 V. The THDP0100 supports bandwidths up to 100 MHz and slew rates up to 2500 V/ns at 1/1000 gain. These probes are supplied with two sizes of hook tips and have an overrange visual and audible indicator which warns the user when they are exceeding the linear range of the probe.

The THDP0200 and P5205A are active differential probes that are capable of safely measuring differential voltages up to ±1500 V. The probes are effective in making measurements in IGBT circuits such as motor drives and power converters. The THDP0200 supports bandwidths up to 200 MHz and slew rates up to 275 V/ns at 1/250 gain.

THDP0100高压差分探头are designed for medium-voltage applications with differential requirements up to ±750 V. These probes have lower attenuation ranges and offer better signal-to-noise ratio.

The P5200A can be used with any oscilloscope and enables users to safely make measurements of floating circuits with their oscilloscope grounded. The P5200A Active Differential Probe converts floating signals to low-voltage ground-referenced signals that can be displayed safely and easily on any ground-referenced oscilloscope.

Warning: For safe operation, do not use the P5200A High-voltage Differential Probe with oscilloscopes that have floating inputs (isolated inputs), such as the Tektronix TPS2000 and THS3000 Series oscilloscopes. The P5200A High-voltage Differential Probe requires an oscilloscope or other measurement instrument with grounded inputs.

Connectivity options

The TMDP and THDP Series probes are equipped with the TekVPI®interface which allows smart communication between the oscilloscope and probe. Pushing the probe menu button will launch the probe control menu on the oscilloscope display providing access to all relevant probe settings and controls. These probes are designed to operate on TekVPI®oscilloscopes without requiring the TPA-BNC adapter.

The P52xxA Series probes are equipped with the Tektronix TekProbe®interface which communicates scale information to the oscilloscope. Direct connections can be made to oscilloscopes configured with the TekProbe®interface or to any oscilloscope when used with the 1103 TekProbe®Power Supply.