技术文章

小鼠白介素10(IL-10)ELISA试剂盒说明书

厦门慧嘉生物科技有限公司 >> 进入商铺

2011/8/11 17:58:35

小鼠白介素10(IL-10)ELISA试剂盒

 

厦门慧嘉生物经营ELISA试剂盒及抗体、细胞因子、生化试剂、耗材等生物试剂产品。诚信经营,价格实惠,服务周到,质量有保证。:   : http://www.biohj.com/download.aspx(说明书下载) 该说明书是PDF格式转化的,固排版有所变化, 欢迎老师或或索取原版说明书

----------------------- Page 1-----------------------

       Mouse Interleukin 10(IL-10)

                              ELISA Kit

 

                        Catalog No. CSB-E04594m

 

                                       (96T)

 

    This immunoassay kit allows for the in vitro quantitative determination of mouse

 

    IL-10 concentrations in serum, plasma and other biological fluids.

 

    Expiration date     six months from the date of manufacture

 

    FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.

 

                                           1


----------------------- Page 2-----------------------

INTRODUCTION

 

Interleukin-10   (IL-10), also   known   as human  cytokine   synthesis   inhibitory

 

factor (CSIF), is an anti-inflammatory cytokine. This cytokine is produced

 

primarily     by   monocytes       and    to  a  lesser    extent    by   lymphocytes.      This

 

cytokine   has   pleiotropic   effects   in   immunoregulation   and   inflammation.   It

 

down-regulates   the   expression   of   Th1   cytokines,   MHC  class   II   antigens,

 

and   costimulatory   molecules   on   macrophages.   It   also   enhances   B   cell

 

survival,   proliferation,   and   antibody   production.   This   cytokine   can   block

 

NF-κB activity, and is involved in the regulation of the JAK-STAT signaling

 

pathway. Knockout studies in mice suggested the function of this cytokine

 

as an essential immunoregulator in the intestinal tract. A study in mice has

 

shown that interleukin-10 is also produced by mast cells, counteracting the

 

inflammatory effect that these cells have at the site of an allergic reaction.

 

PRINCIPLE OF THE ASSAY

 

The     microtiter    plate   provided    in  this   kit  has   been   pre-coated       with   an

 

antibody   specific   to   IL-10   Standards   or   samples   are   then   added   to   the

 

appropriate       microtiter    plate    wells   with    a   biotin-conjugated       polyclonal

 

antibody       preparation      specific     for   IL-10    and     Avidin    conjugated       to

 

Horseradish        Peroxidase      (HRP)     is  added     to  each    microplate     well   and

 

incubated. Then a TMB (3,3'5, 5' tetramethyl-benzidine) substrate solution

 

is added to each well. Only those wells that contain IL-10, biotin-conjugated

 

antibody and enzyme-conjugated Avidin will exhibit a change in color. The

 

enzyme-substrate reaction is terminated by the addition of a sulphuric acid

 

solution   and   the   color   change   is   measured   spectrophotometrically   at   a

 

wavelength of 450 nm ± 2 nm. The concentration of IL-10 in the samples is

 

then   determined   by   comparing   the   O.D.   of   the   samples   to   the   standard

 

curve.

 

                                                2


----------------------- Page 3-----------------------

DETECTION RANGE

 

3.2   pg/ml-200    pg/ml.   The   standard   curve   concentrations     used   for  the

 

ELISA’s   were   200   pg/ml,   100   pg/ml,   50   pg/ml,   25   pg/ml,   12.5   pg/ml,6.2

 

pg/ml,3.2 pg/ml.

 

SPECIFICITY

 

This assay recognizes recombinant and natural mouse IL-10 No significant

 

cross-reactivity or interference was observed.

 

SENSITIVITY

 

The   minimum   detectable   dose   of   mouse   IL-10   is   typically   less   than   0.8

 

pg/ml.

 

The sensitivity of this assay, or Lower Limit of Detection (LLD) was defined

 

as the lowest protein concentration that could be differentiated from zero.

 

MATERIALS PROVIDED

 

               Reagent                                      Quantity

 

               Assay plate                                       1

 

               Standard                                          2

 

               Sample Diluent                               1 x 20 ml

 

               Biotin-antibody Diluent                      1 x 10 ml

 

               HRP-avidin Diluent                           1 x 10 ml

 

               Biotin-antibody                              1 x 120l

 

               HRP-avidin                                   1 x 120l

 

                                                            1 x 20 ml

               Wash Buffer

                                                         (25×concentrate)

 

               TMB Substrate                                1 x 10 ml

 

               Stop Solution                                1 x 10 ml

 

                                           3


----------------------- Page 4-----------------------

STORAGE

 

1.   Unopened   test   kits   should   be   stored   at   2-8°C   upon   receipt   and   the

 

    microtiter plate should be kept in a sealed bag. The test kit may be used

 

    throughout      the   expiration    date   of   the  kit,  provided    it  is  stored   as

 

    prescribed above. Refer to the package label for the expiration date.

 

2.   Opened test plate should be stored at 2-8°C in the aluminum foil bag

 

    with desiccants to minimize exposure to damp air. The kits will remain

 

    stable until the expiring date shown, provided it is stored as prescribed

 

    above.

 

3.   A   microtiter   plate   reader   with   a   bandwidth   of   10   nm   or   less   and   an

 

    optical   density   range   of   0-3   OD   or   greater   at   450nm   wavelength   is

 

    acceptable for use in absorbance measurement.

 

REAGENT PREPARATION

 

Bring all reagents to room temperature before use.

 

1.   Wash Buffer          If crystals have formed in the concentrate, warm up to

 

     room   temperature   and   mix   gently   until   the   crystals  have   compley

 

     dissolved. Dilute 20 ml of Wash Buffer Concentrate  into deionized or

 

     distilled water to prepare 500 ml of Wash Buffer.

 

2.   Standard        Centrifuge   the   standard   vial   at   6000-10000rpm   for       30s.

 

     Reconstitute       the Standard       with   1.0   ml   of Sample       Diluent.     This

 

     reconstitution      produces     a   stock   solution    of  200    pg/ml.   Allow    the

 

     standard to sit for a minimum of 15 minutes with gentle agitation prior to

 

     making   serial     dilutions.   The   undiluted    standard     serves   as   the  high

 

     standard (200 pg/ml). The Sample Diluent serves as the zero standard

 

     (0 pg/ml). Prepare fresh for each assay. Use within 4 hours and discard

 

     after use.

 

                                               4


----------------------- Page 5-----------------------

3.   Biotin-antibody         Centrifuge     the  vial  before   opening.    Dilute   to  the

 

     working       concentration        using    Biotin-antibody          Diluent(1:100),

 

     respectively.

 

4.   HRP-avidin        Centrifuge the vial before opening. Dilute to the working

 

     concentration using HRP-avidin Diluent(1:100), respectively.

 

Precaution: The Stop Solution provided with this kit is an acid solution. Wear

 

             eye, hand, face, and clothing protection when using this material.

 

OTHER SUPPLIES REQUIRED

 

      Microplate   reader   capable   of   measuring   absorbance  at   450   nm,   with

 

     the correction wavelength set at 540 nm or 570 nm.

 

      Pipettes and pipette tips.

 

      Deionized or distilled water.

 

      Squirt bottle, manifold dispenser, or automated microplate washer.

 

      An incubator which can provide stable incubation conditions up to

 

     37°C±0.5°C.

 

SAMPLE COLLECTION AND STORAGE

 

     Serum      Use a serum separator tube (SST) and allow samples to clot

 

     for 30 minutes before centrifugation for 15 minutes at 1000 g. Remove

 

     serum and assay immediay or aliquot and store samples at -20°C.

 

     Centrifuge   the   sample   again   after   thawing   before   the   assay.   Avoid

 

     repeated freeze-thaw cycles.

 

     Plasma       Collect     plasma     using   citrate,  EDTA,     or   heparin    as   an

 

     anticoagulant. Centrifuge for 15 minutes at 1000 g within 30 minutes of

 

     collection.   Assay   immediay   or   aliquot   and   store  samples   at   -20°C.

 

     Centrifuge   the   sample   again   after   thawing   before   the   assay.   Avoid

 

     repeated freeze-thaw cycles.

 

Note: Grossly hemolyzed samples are not suitable for use in this assay.

 

                                              5


----------------------- Page 6-----------------------

ASSAY PROCEDURE

 

Bring     all  reagents    and    samples     to  room     temperature    before     use.   It  is

recommended that all samples, standards, and controls be assayed in duplicate.

All   the   reagents   should   be   added   directly   to   the   liquid   level   in   the   well.   The

pipette should avoid contacting the inner wall of the well.

 

 1.   Add   100l   of   Standard,   Blank,   or   Sample   per   well.   Cover   with   the

 

     adhesive strip. Incubate for 2 hours at 37°C.

 

 2.   Remove the liquid of each well, don’t wash.

 

 3.   Add 100l of Biotin-antibody working solution to each well. Incubate

 

     for   1   hour   at  37°C.   Biotin-antibody   working           solution    may    appear

 

     cloudy.   Warm   up   to   room   temperature   and   mix   gently  until   solution

 

     appears uniform.

 

 4.   Aspirate each well and wash, repeating the process three times for a

 

     total of three washes. Wash: Fill each well with Wash Buffer (200l) and

 

     let   it  stand for  2  minutes, then   remove   the   liquid  by flicking  the  plate

 

     over a sink. The remaining drops are removed by patting the plate on a

 

     paper   towel.   Complete   removal   of   liquid   at   each   step   is   essential   to

 

     good performance.

 

 5.   Add   100l   of  HRP-avidin   working   solution   to   each   well.   Cover   the

 

     microtiter plate with a new adhesive strip. Incubate for 1 hour at 37°C.

 

 6.   Repeat the aspiration and wash three times as step 4.

 

 7.   Add 90l of TMB Substrate to each well. Incubate for 10-30 minutes at

 

     37°C.     Keeping     the   plate   away     from   drafts    and   other   temperature

 

     fluctuations in the dark.

 

 8.   Add    50l    of  Stop    Solution   to     each    well  when     the  first   four  wells

 

     containing the highest concentration of standards develop obvious blue

 

     color. If color change does not appear uniform, gently tap the plate to

 

     ensure thorough mixing.

 

 9.   Determine the optical density of each well within 30 minutes, using a

 

     microplate reader set to 450 nm.

 

                                                6


----------------------- Page 7-----------------------

CALCULATION OF RESULTS

 

Using   the   professional   soft   "Curve   Exert   1.3"   to   make   a   standard   curve   is

 

recommended, which can be downloaded from our web.

 

Average the duplicate readings for each standard, control, and sample and

 

subtract the average zero standard optical density. Create a standard curve

 

by reducing the data using computer software capable of generating a four

 

parameter logistic (4-PL) curve-fit. As an alternative, construct a standard

 

curve   by   plotting   the   mean   absorbance   for   each   standard   on   the   y-axis

 

against the concentration on the x-axis and draw a best fit curve through the

 

points on the graph. The data may be linearized by plotting the log of the

 

IL-10 concentrations versus the log of the O.D. and the best fit line can be

 

determined       by    regression     analysis.    This   procedure      will  produce      an

 

adequate but less precise fit of the data. If samples have been diluted, the

 

concentration       read   from   the   standard    curve    must   be  multiplied     by  the

 

dilution factor.

 

LIMITATIONS OF THE PROCEDURE

 

      The kit should not be used beyond the expiration date on the kit label.

 

      Do not mix or substitute reagents with those from other lots or sources.

 

      It is important that the Standard Diluent selected for the standard curve

 

     be consistent with the samples being assayed.

 

      If samples generate values higher than the highest standard, dilute the

 

     samples with the appropriate Standard Diluent and repeat the assay.

 

      Any    variation    in  Standard      Diluent,    operator,    pipetting    technique,

 

     washing   technique,   incubation   time   or   temperature,  and   kit   age   can

 

     cause variation in binding.

 

                                               7


----------------------- Page 8-----------------------

      This assay is designed to eliminate interference by soluble receptors,

 

     binding proteins, and other factors present in biological samples. Until

 

     all   factors   have   been   tested   in   the   Immunoassay,   the   possibility   of

 

     interference cannot be excluded.

 

TECHNICAL HINTS

 

      Centrifuge vials before opening to collect contents.

 

      When mixing or reconstituting protein solutions, always avoid foaming.

 

      To avoid cross-contamination, change pipette tips between additions of

 

     each standard level, between sample additions, and between reagent

 

     additions. Also, use separate reservoirs for each reagent.

 

      When   using   an   automated   plate   washer,   adding   a   30  second   soak

 

     period   following   the   addition   of   wash   buffer,   and/or   rotating   the   plate

 

     180 degrees between wash steps may improve assay precision.

 

      To   ensure   accurate   results,   proper   adhesion   of   plate   sealers   during

 

     incubation steps is necessary.

 

      Substrate Solution should remain colorless or light blue until added to

 

     the    plate.   Keep    Substrate     Solution    protected     from    light.  Substrate

 

     Solution   should   change   from   colorless   or   light   blue   to   gradations   of

 

     blue.

 

      Stop Solution should be added to the plate in the  same order as the

 

     Substrate Solution. The color developed in the wells will turn from blue

 

     to   yellow   upon   addition   of   the   Stop   Solution. Wells  that   are   green   in

 

     color indicate that the Stop Solution has not mixed thoroughly with the

 

     Substrate Solution.

 

                                                8


----------------------- Page 9-----------------------

                 小鼠小鼠白介素白介素10(IL-10)酶联免疫分析酶联免疫分析

                 小鼠小鼠白介素白介素                   酶联免疫分析酶联免疫分析

 

                           试剂盒使用说明书试剂盒使用说明书

                           试剂盒使用说明书试剂盒使用说明书

 

本试剂盒仅供研究使用本试剂盒仅供研究使用

本试剂盒仅供研究使用本试剂盒仅供研究使用

 

产品编号产品编号::CSB-E04594m

产品编号产品编号::

 

检测范围检测范围::3.2 pg/ml-200 pg/ml,

检测范围检测范围::

 

zui低检测zui低检测限:限:0.8 pg/ml

zui低检测zui低检测限限::

 

特异性特异性::本试剂盒可同时检测天然或重组的小鼠 IL-10,且与其他相关蛋

特异性特异性::

白无交叉反应。

 

有效期有效期::6 个月

有效期有效期::

 

预期应用预期应用::ELISA 法定量测定小鼠血清、血浆、细胞培养上清或其它相关

预期应用预期应用::

生物液体中IL-10 含量。

 

说明说明

说明说明

 

1   试剂盒保存:未开封的试剂盒应储存于2-8℃;开封后的酶标板应与干燥

    剂一起储存于铝箔袋中置于2-8℃保存。仅在此出储存条件下,产品在有

    效期内可正常使用。

2   浓洗涤液低温保存会有盐析出,稀释时可在水浴中加温助溶。

3   中、英文说明书可能会有不一致之处,请以英文说明书为准。

4   刚开启的酶联板孔中可能会含有少许水样物质,此为正常现象,不会对实

    验结果造成任何影响。

 

实验原理实验原理

实验原理实验原理

 

     用纯化的抗体包被微孔板,制成固相载体,往包被抗IL-10 抗体的微孔中

依次加入标本或标准品、生物素化的抗IL-10 抗体、HRP 标记的亲和素,经

过*洗涤后用底物TMB 显色。TMB 在过氧化物酶的催化下转化成蓝色,

并在酸的作用下转化成zui终的黄色。颜色的深浅和样品中的IL-10 呈正相关。

用酶标仪在450nm 波长下测定吸光度 (OD 值),计算样品浓度。

 

                                      9


----------------------- Page 10-----------------------

试剂盒组成及试剂配制试剂盒组成及试剂配制

试剂盒组成及试剂配制试剂盒组成及试剂配制

 

1.  酶联板酶联板(Assay plate ):                                      一块(96 孔)。

    酶联板酶联板

 

2.  标准品标准品(Standard):                                          2 瓶(冻干品)。

    标准品标准品

 

3.  样品稀释液样品稀释液(Sample Diluent):                                  1×20ml/瓶。

    样品稀释液样品稀释液

 

4.  生物素标记抗体稀释液生物素标记抗体稀释液((Biotin-antibody Diluent):)             1×10ml/瓶。

    生物素标记抗体稀释液生物素标记抗体稀释液((                            ))

 

5.  辣根过氧化物酶标记亲和素稀释液辣根过氧化物酶标记亲和素稀释液((HRP-avidin Diluent)) 1×10ml/瓶。

    辣根过氧化物酶标记亲和素稀释液辣根过氧化物酶标记亲和素稀释液((                         ))

 

6.  生物素标生物素标记抗体记抗体((Biotin-antibody):)                 1×120l/瓶(1:100)。

    生物素标生物素标记抗体记抗体((                  ))

 

7.  辣根过氧化物酶标记亲和素辣根过氧化物酶标记亲和素((HRP-avidin):)            1×120l/瓶(1:100)。

    辣根过氧化物酶标记亲和素辣根过氧化物酶标记亲和素((                ))

 

8.  底物溶液底物溶液((TMB Substrate):)                                   1×10ml/瓶。

    底物溶液底物溶液((                 ))

 

9.  浓洗涤液浓洗涤液((Wash Buffer )) 1×20ml/瓶,使用时每瓶用蒸馏水稀释25 倍。

    浓洗涤液浓洗涤液((               ))

 

10. 终止液终止液((Stop Solution):)                                     1×10ml/瓶。

    终止液终止液((                ))

 

需要而未提供的试剂和器材需要而未提供的试剂和器材

需要而未提供的试剂和器材需要而未提供的试剂和器材

 

1.  标准规格酶标仪

 

2.  高速离心机

 

3.  电热恒温培养箱

 

4.  干净的试管和Eppendof 管

 

5.  系列可调节移液器及吸头,一次检测样品较多时,用多通道移液器

 

6.  蒸馏水,容量瓶等

 

标本的采集及保存标本的采集及保存

标本的采集及保存标本的采集及保存

 

1.  血清:全血标本请于室温放置2 小时或4℃过夜后于1000g 离心20 分钟,

 

    取上清即可立即检测;或进行分装,并将标本放于-20℃或-80℃保存,但

 

    应避免反复冻融。解冻后的样品应再次离心,然后检测。

 

2.  血浆:可用EDTA 或肝素作为抗凝剂,标本采集后30 分钟内于2   -   8°C

 

    1000 g 离心15 分钟,取上清即可立即检测;或进行分装,并将标本放于

 

    -20℃或-80℃保存,但应避免反复冻融。解冻后的样品应再次离心,然后

 

    检测。

 

注:注:标本溶血会影响zui后检测结果标本溶血会影响zui后检测结果,因此溶血标本不宜进行此项检测,因此溶血标本不宜进行此项检测。。

注注::标本溶血会影响zui后检测结果标本溶血会影响zui后检测结果,,因此溶血标本不宜进行此项检测因此溶血标本不宜进行此项检测。。

 

                                      10


----------------------- Page 11-----------------------

标准品的稀释原则标准品的稀释原则::2 瓶,使用前于6000-10000rpm 离心30 秒。每

标准品的稀释原则标准品的稀释原则::

瓶临用前以样品稀释液稀释至1ml,盖好后静置10 分钟以上,然后反复颠倒

/搓动以助溶解,其浓度为200 pg/ml,做系列倍比稀释后,分别稀释200 pg/ml,

100 pg/ml, 50 pg/ml, 25 pg/ml, 12.5 pg/ml,6.2 pg/ml,3.2 pg/ml,样品稀释液

直接作为标准浓度0   pg/ml,临用前15 分钟内配制,用完丢弃,下次检测使

用新鲜配置的标准品。

 

如配制100   pg/ml 标准品:取0.5ml  (不要少于0.5ml)200   pg/ml 的上述标

准品加入含0.5ml 样品稀释液的Eppendorf 管中,混匀即可,其余浓度以此

类推。

 

生物素标记抗体的稀释原则生物素标记抗体的稀释原则::

生物素标记抗体的稀释原则生物素标记抗体的稀释原则::

 

打开瓶盖前请离心,收集瓶壁上的溶液。临用前以生物素标记抗体稀释液稀

释,稀释前根据预先计算好的每次实验所需的总量配制 (每孔100l),实际

配制时应多配制0.1-0.2ml。如10l 生物素标记抗体加990l 生物素标记抗体

稀释液的比例配制,轻轻混匀,在使用前一小时内配制。

 

辣根过氧化物酶标记亲和素的稀释原则辣根过氧化物酶标记亲和素的稀释原则::

辣根过氧化物酶标记亲和素的稀释原则辣根过氧化物酶标记亲和素的稀释原则::

 

打开瓶盖前请离心,收集瓶壁上的溶液。临用前以辣根过氧化物酶标记亲和

素稀释液稀释,稀释前根据预先计算好的每次实验所需的总量配制(每孔

100l),实际配制时应多配制0.1-0.2ml。如10l 辣根过氧化物酶标记亲和素

加990l 辣根过氧化物酶标记亲和素稀释液的比例配制,轻轻混匀,在使用

前一小时内配制。

 

操作步骤操作步骤

操作步骤操作步骤

 

实验开始前,请提前配置好所有试剂,试剂或样品稀释时,均需混匀,混匀

时尽量避免起泡。每次检测都应该做标准曲线。如样品浓度过高时,用样品

稀释液进行稀释,以使样品符合试剂盒的检测范围。加样时,枪头应直接对

准液面,切勿沿孔壁加样。

1   加样:分别设空白孔、标准孔、待测样品孔。空白孔加样品稀释液100l,

    余孔分别加标准品或待测样品100l,注意不要有气泡,加样将样品加于

    酶标板孔底部,尽量不触及孔壁,轻轻晃动混匀,酶标板加上盖或覆膜,

    37℃反应120 分钟。

    为保证实验结果有效性,每次实验请使用新的标准品溶液。

 

                                      11


----------------------- Page 12-----------------------

2   弃去液体,甩干,不用洗涤。每孔加生物素标记抗体工作液 100l (取1l

    生物素标记抗体加99l 生物素标记抗体稀释液的比例配制,轻轻混匀,

    在使用前一小时内配制),37℃,60 分钟。

3   温育60 分钟后,弃去孔内液体,甩干,洗板3 次,每次浸泡1-2 分钟,

    200l/每孔,甩干。

4   每孔加辣根过氧化物酶标记亲和素工作液 (同生物素标记抗体工作液)

    100l,37℃,60 分钟。

5   温育60 分钟后,弃去孔内液体,甩干,洗板5 次,每次浸泡1-2 分钟,

    200l/每孔,甩干。

6   依序每孔加底物溶液90l,37℃避光显色((30 分钟内,此时肉眼可见标

                                                ((

    准品的前3-4 孔有明显的梯度蓝色,后3-4 孔显色不明显,即可终止)。

7   依序每孔加终止溶液50l,终止反应(此时蓝色立转黄色)。终止液的加

    入顺序应尽量与底物液的加入顺序相同。为了保证实验结果的准确性,底

    物反应时间到后应尽快加入终止液。

8   用酶联仪在450nm 波长依序测量各孔的光密度 (OD 值)。 在加终止液

    后15 分钟以内进行检测。

 

实验备注实验备注

实验备注实验备注

 

1.  用户在初次使用试剂盒时,应将各种试剂管离心数分钟,以便试剂集中到

    管底。

2.  每次实验留一孔作为空白调零孔,该孔不加任何试剂,只是zui后加底物溶

    液及终止液。测量时先用此孔调OD 值至零。

3.  为防止样品蒸发,试验时将反应板放于铺有湿布的密闭盒内,酶标板加上

    盖或覆膜。

4.  未使用完的酶标板或者试剂,请于2-8℃保存。标准品、生物素标记抗体

    工作液、辣根过氧化物酶标记亲和素工作液请依据所需的量配置使用。请

    勿重复使用已稀释过的标准品、生物素标记抗体工作液、或辣根过氧化物

    酶标记亲和素工作液。

5.  建议检测样品时均设双孔测定,以保证检测结果的准确性。

 

洗板方法洗板方法

洗板方法洗板方法

手工洗板方法:吸去(不可触及板壁)或甩掉酶标板内的液体;在实验台上

铺垫几层吸水纸,酶标板朝下用力拍几次;将推荐的洗涤缓冲液至少0.3ml

注入孔内,浸泡1-2 分钟。根据需要,重复此过程数次。

 自动洗板:如果有自动洗板机,应在熟练使用后再用到正式实验过程中。

 

                                      12


----------------------- Page 13-----------------------

计算计算

计算计算

 

请从我们的下载专业软件"Curve Exert 1.3",并根据提示制作标准曲线。

以标准物的浓度为横坐标(对数坐标),OD 值为纵坐标(普通坐标),在半对

数坐标纸上绘出标准曲线,根据样品的OD 值由标准曲线查出相应的浓度;

再乘以稀释倍数;或用标准物的浓度与OD 值计算出标准曲线的直线回归方

程式,将样品的OD 值代入方程式,计算出样品浓度,再乘以稀释倍数,即

为样品的实际浓度。

 

注意事项注意事项

注意事项注意事项

 

1.  本操作说明适用于本操作说明适用于48T 试剂盒试剂盒,, 48T 试剂盒中试剂盒中酶联板酶联板、标准品、标准品、生物素、生物素

    本操作说明适用于本操作说明适用于        试剂盒试剂盒,,      试剂盒中试剂盒中酶联板酶联板、、标准品标准品、、生物素生物素

    标记抗体及辣根过氧化物酶标记亲和素减半标记抗体及辣根过氧化物酶标记亲和素减半。。

    标记抗体及辣根过氧化物酶标记亲和素减半标记抗体及辣根过氧化物酶标记亲和素减半。。

2.  当混合蛋白溶液时应尽量轻缓,避免起泡。

3.  洗涤过程非常重要,不充分的洗涤易造成假阳性。

4.  一次加样时间控制在5 分钟内,如标本数量多,推荐使用排枪加样。

5.  请每次测定的同时做标准曲线,做复孔。

6.  如标本中待测物质含量过高,请先稀释后再测定,计算时请zui后乘以稀释

    倍数。

7.  在配制标准品、检测溶液工作液时,请以相应的稀释液配制,不能混淆。

8.  底物请避光保存。

9.  不要用其它生产厂家的试剂替换试剂盒中的试剂。

 

                                      13


----------------------- Page 14-----------------------

Notes

 

                                                                            14


----------------------- Page 15-----------------------

                                                                            15


----------------------- Page 16-----------------------

                                                                            16


----------------------- Page 17-----------------------

                                                                            17


----------------------- Page 18-----------------------

                                                                            18


----------------------- Page 19-----------------------

                                                                            19


----------------------- Page 20-----------------------

 

相关产品

猜你喜欢

当前客户在线交流已关闭
请电话联系他 :