ZJC-150KV计算机控制塑料薄膜介电击穿/绝缘材料耐电压强度测试仪
ZJC-150KV computer controlled voltage breakdown tester
一、应用范围:
One, application range:
ZJC-150KV主要适用于固体绝缘材料工频电压击穿的测试如电线套管、树脂和胶、浸渍纤维制品、云母及其制品、塑料、薄膜复合制品、陶瓷和玻璃等介质在工频电压或直流电压下击穿强度和耐电压时间的测试;击穿电压、介电强度测试仪器采用计算机控制,可对介电强度试验过程中的各种数据进行快速、准确的采集、处理,并可存取、显示、打印。
ZJC-150KV is mainly applied to the test of power frequency voltage breakdown of solid insulating materials such as wire sleeve, resin and rubber, impregnated fiber products, mica and its products, plastic, film composite products, ceramics and glass, etc., which can be used in the test of power frequency voltage or DC voltage.
二、满足标准:
Two, meet the standard:
1、GB1408.1-2006《绝缘材料电气强度试验方法》
1, GB1408.1-2006 "electrical strength test method for insulation materials"
2、GB1408.2-2006《绝缘材料电气强度试验方法 ,第2部分:对应用直流电压试验的附加要求》
2, "GB1408.2-2006 electrical strength of insulating materials test methods, part second: Additional requirements for the application of DC voltage test"
3、GB/T1695-2005《硫化橡胶工频击穿电压强度和耐电压的测定方法》
3, GB/T1695-2005 "vulcanized rubber power frequency breakdown voltage strength and resistance to voltage measurement method"
4、GB/T3333电缆纸工频电压击穿试验方法
4, GB/T3333 cable paper power frequency voltage breakdown test method
5、HG/T 3330绝缘漆漆膜击穿强度测定法
5, HG/T 3330 insulating paint film breakdown strength test method
6、GB12656 电容器纸工频电压击穿试验方法
6, GB12656 capacitor paper power frequency voltage breakdown test method
7、ASTM D149《固体电绝缘材料工业电源频率下的介电击穿电压和介电强度的试验方法》
7, D149 ASTM "test method for dielectric breakdown voltage and dielectric strength of the dielectric breakdown voltage at the industrial power supply frequency of the solid electrical insulation material"
三、主要技术指标:
Three, the main technical indicators:
1、输入电压: AC220V 50Hz ;
1, input voltage: AC220V 50Hz;
2、输出电压: AC:0~150kV; DC:0~150kV;
2, output voltage: AC:0 ~ 150kV ~ 150kV; DC:0;
3、输出功率:15kVA;
3, output power: 15kVA;
5、测量误差: ≤ 5%;
5, the measurement error is less than 5%;
6、升压速率:1kV/s~3kV/s;
6, boost rate: 1kV/s ~ 3kV/s;
7、耐压时间: 0~4H;
7, pressure time: 0 ~ 4H;
8、漏电流 : 1~30 mA可由计算机软件自由进行设定;
8, the leakage current: 1 ~ 30 mA can be set free by computer software;
9、电源:交流220V±10%的单相交流电压和50Hz±1%的频率;
9, power: AC 220V + 10% single-phase AC voltage and frequency of 50Hz 1%;
10、试验环境温度: 15 ~ 30℃,相对湿度:0~85%能够稳定运行;
10, test environment temperature: 15 ~ 30, relative humidity: 0 ~ 85% can be stable operation;
11、外形尺寸长×宽×高:1980mm×1220 mm×1750mm;
11, the shape of the size of a long x x x x x x x x x x x x x x x x x x x x x x x x;
12、设备自重:350Kg;
12, equipment weight: 350Kg;
13、接地要求仪器需要单独接地,接地附合国家标准要求,金属棒深埋地下至少要1.5米以下;
13, grounding requires that the instrument need to be alone, the ground attached to the national standard requirements, metal rods deep underground at least 1.5 meters below;
四、性能特点:
Four, performance characteristics:
1、高压变压器主要产生试样所需的直流电压;
1, high voltage transformer mainly produces the sample required for the DC voltage;
2、调压器用于调节升压变压器输入端电压以产生高压所需的输入电压;
2, the regulator is used to adjust the input voltage of the boost transformer to generate the input voltage required by the high voltage;
3、电压测量主要是从高压变压器测量端测量,高压变压器测量端和高压端是线性的;
3, voltage measurement is mainly from the measurement of high voltage transformer, high voltage transformer measurement and high voltage side is linear;
4、放电系统在试验做完以后放电,以免产生放电对人身的危害;
4, the discharge system in the test after the discharge, so as to avoid the discharge of human body damage;
五、计算机系统及软件包:
Five, computer system and software package:
1、试验软件是我公司研发的功能强大、操作简单、显示直观的试验软件系统。
1, testing software is the latest research and development of our company is powerful, simple operation, display and intuitive test software system.
2、采用计算机控制通过人机对话方式,完成对绝缘介质工频电压击穿,工频耐压试验。
2, the use of computer control through the man-machine dialogue, the completion of the insulation medium power frequency voltage breakdown, power frequency withstand voltage test.
3、试验过程中可动态绘制出试验曲线,试验的曲线可以多种颜色叠加对比,局部放大,曲线上任意一段可进行区域放大分析;
3, the test process can be drawn from the test curve, the curve can be a variety of color overlay contrast, local amplification, a section of the curve can be amplified in any region;
4、可对试验数据进行编辑修改,灵活适用;
4, can be used to edit and modify the test data, flexible application;
5、试验条件及测试结果等数据可自动存储;
5, test conditions and test results and other data can be automatically stored;
6、试验报告格式灵活可变,适用于不同用户的不同需求;
6, the test report format is flexible, suitable for different needs of different users;
7、可对一组试验中曲线数据的有效与否进行人为选定;
7, can be used for a set of test data in the effective or not for the selection of human;
8、试验结果数据可导入EXECL,WORD文档编辑;
8, test results data can be imported into WORD, EXECL document editing;
六、安全说明:
Six, safety instructions:
1、介电强度仪器设备要安装单独的保护地线。接保护地线,主要是减少试样击穿时对周围产生的较强的电磁干扰。也可避免控制计算机失控。
1, the dielectric strength of the instrument equipment to install a separate ground wire. Earth wire, mainly to reduce the sample breakdown of the strong electromagnetic interference around the. Can also avoid control computer control.
2、该介电强度仪器设备的电路设有多项保护措施,主要有:过流保护、失压保护、漏电保护、短路保护、直流试验放电报警等。
2, the dielectric strength of the instrument and equipment of the circuit with a number of protective measures, mainly include: over current protection, loss of voltage protection, leakage protection, short circuit protection, DC test discharge alarm, etc..
3、安全保护:
3, safety protection:
(1)门限位保护:不关门,即使通电点实验开始,设备无任何反应,软件有:安全门未关闭提示。
(1) the threshold level protection: do not close the door, even though the power point experiment, the device has no any response, the software is: the safety door is not closed.
(2)电压归零保护:如果在实验过程中,突然断电,下次开机后,会自动回到零位,保证初始电压为在零位
(2) the voltage to zero protection: if the experimental process, the power, the next boot, will automatically return to zero, to ensure that the initial voltage is zero
(3)终止电压保护:可以通过软件设定终止电压,保证在升压过程中如果出现异常升到的电压后,自动终止并归零
(3) termination voltage protection: it can be set by the software to terminate the voltage, to ensure that in the process of the rise in the event of abnormal rise to the specified voltage, automatic termination and return to zero
(4)高压机械限位:如果软件系统失去控制,电压继续往上升,到高压限位后自动归零如果在击穿后未判停,通过过流保护器采集数据保证电压自动归零有高压指示灯,通过观察指示灯的状态来判断实在升压还是在零位如果长时间做实验,为保证设备的良好运行,设备留有变压器排气口,保证变压器的良好运行,增长使用寿命如果在实验中,试样有异味或者出现燃烧和冒烟现象,可以通过排风系统进行排除。
(4) high voltage mechanical limit: if the software system is out of control, the voltage continues to rise, to the high voltage limit automatically after the breakdown of the non stop, through the flow protection device to collect data to ensure that the voltage automatically to zero with high pressure indicator, through the observation of the status of the test, to ensure the good operation of equipment, equipment to keep a good run, the growth of the use of life in the experiment, if the sample has a smell or smoke phenomenon, can be excluded through exhaust system.
(5)独立接地保护
(5) independent grounding protection
(6)短路保护
(6) short circuit protection
(7)软件误操作保护
(7) software error protection
(8)漏电保护
(8) leakage protection
(9)实验结束放电保护
(9) the end of the experiment is to end the discharge protection