氯乙烯CH2=CHCl气体检知管132SB VINYL CHLORIDE 5-500ppm
Tube 32SB
1. PERFORMANCE
1) Measuring range :5-500 ppm
Number of pump strokes 1(100ml)
2) Sampling time :1.5 minutes/1 pump stroke
3) Detectable limit :2 ppm
4) Shelf life :1.5 years(Necessary to store in a refrigerated place ; 0~10℃)
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :White→Reddish orange
2. RELATIVE STANDARD DEVIATION
RSD-low:15% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
By decomposing with an Oxidizer, Chlorine is produced. This Chlorine reacts with o-Toluidine and red
Orthoquinone is produced.
CH2=CHCI+PbO2→CI2
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
PRETREAT TUBE DETECTOR TUBE
500
400
300
200
100
50
20
10
5
ppm
(Gray)
132SB VINYL CHLORIDE
Substance Interference ppm Coexistence
Hydrogen chloride Similar stain is produced. Higher readings are given.
Chlorine 〃〃
Ohytdhreorc haarbloognesns or Halogenated 〃〃
TEMPERATURE CORRECTION TABLE
Scale True Concentration (ppm)
Readings 0℃ 5℃ 10℃ 15℃ 20℃ 30℃ 40℃
(ppm) (32°F) (41°F) (50°F) (59°F) (68°F) (86°F) (104°F)
500 - - 820 600 500 430 400
400 - 850 610 470 400 340 310
300 - 640 440 350 300 250 230
200 750 400 300 230 200 160 140
100 340 200 150 120 100 80 70
50 150 100 70 60 50 40 30
20 50 40 30 25 20 15 10
10 25 20 15 10 10 10 5
5 15 10 10 5 5 5 5
NH2+CI2
CH3 CH3
H2N CIN=� =� =�NCI
CH3 CH3