EEG CMRR tester CMRR3.0E, combining accuracy, convenience and test automation all together.
Compliant with IEC80601-2-26:2019 and applicable for R&D, compliance and production run testing.
SPECIFICATION
ItemsOptionsSpecifications
Supply voltage0.5 / 2.828 / 20 / 2 / (70.71) (Vrms)±1%*
CM point voltage1.414 / 10 / 1 / (35.355) (Vrms)±1% (0.25Vrms ±2%)
Frequency50 / 60 / 100 / 120 (Hz)±1%
Electrode with ImpedanceChange electrode via the touch screen of CMRR 3.0ENone / RA / LA / LL / V1 / V2 / V3 / V4 / V5 / V6 / ALL
Electrode without ImpedanceChange electrode via the touch screen of CMRR 3.0ERA / LA / LL / V1 / V2 / V3 / V4 / V5 / V6
Imbalance impedance, R10kΩ10kΩ ±1%
Imbalance impedance, C47nF47nF ± 5%
DC offsetInternal battery powered, can be added on RA / LA / LL / V1~V6150mV ± 1%
Up to 40hrs with intermittent use
100pF capacitorUse 11:1 (110MΩ:10MΩ) voltage divider to measure indirectly100pF ± 5%
EnvironmentIntended for normal laboratory environment. The selection of critical components is known to be stable in the range shown. The 110MΩ divider may be affected by high humidity in excess of 85%
15 - 30°C
10 - 75% RH
*Require to measure with the multimeter of higher accuracy.
设备尺寸: 19.8cm(长)*13.5cm(宽)*8cm(高),屏蔽盒尺寸: 20.4cm(长)*14.7cm(宽)*12.1cm(高)
EEG CMRR tester CMRR3.0E