起订量:
55000-XS (衍生化 串联质谱法LC-MS/MS)
中级会员第7年
代理商温馨提示:仅用于科研,不可用于临床治疗
MassChrom® Amino Acids and Acylcarnitines from dried blood - LC-MS/MS
祥生兴业
960 analyses
The Chromsystems reagent kit for the determination of amino acids and acylcarnitines from dried blood spot samples using tandem mass spectrometry is intended for the early diagnosis of amino acid and fatty acid metabolic disorders during newborn screening.
In Germany, newborn screening is regulated by law. Twelve diseases are defined, eight of them to be determined by tandem mass spectrometry. As therapies for only these specific diseases are known, the diagnosis of other diseases is prohibited. Included in these diseases are defects of amino acid metabolism (phenylketonuria and maple syrup urine disease), defects of fatty acid metabolism (Medium and Very Long Chain Acyl-CoA Dehydrogenase Deficiencies) and organoacidurias (isovaleric acidemia). If not detected, these metabolic errors may impair the newborn severely and irreversibly within the first days of its life.
Linearity: | Amino acids: 2000-2400 µmol/l (Acyl-)carnitines: 25-200 µmol/l |
Limit of quantification: | Amino acids: 2-16 µmol/l (Acyl-)carnitines: 0.1-1.6 µmol/l |
Recovery: | Amino acids: 67-90 % (Acyl-)carnitines: 81-97 % |
Intra-assay: | Amino acids: 3-9 % (Acyl-)carnitines: 4-16 % |
Inter-assay: | Amino acids: 6-11 % (Acyl-)carnitines: 7-20 % |
Analysis time: | < 2 min |
Linearity: | Amino acids with SUAC: 250-2500 µmol/l (Acyl-)carnitines: 11-240 µmol/l |
Limit of quantification: | Amino acids with SUAC: 0.5-5 µmol/l (Acyl-)carnitines: 0.05-0.5 µmol/l |
Recovery: | Amino acids with SUAC: 51-94 % (Acyl-)carnitines: 62-89 % |
Intraassay: | Amino acids with SUAC: 4-10 % (Acyl-)carnitines: 4-20 % |
Interassay: | Amino acids with SUAC: 5-9 % (Acyl-)carnitines: 8-28 % |
Analysis time: | < 2 min |
Amino acids: | 22-743 µmol/l |
Free carnitine: | 50-60 µmol/l |
Acylcarnitines: | 0.1-48 µmol/l |
SUAC: | 1.16 µmol/l |
LC-MS/MS (tandem MS). No HPLC column is required.
Injection volume: | 10 µl |
Flow gradient: | 20-600 µl/min |
Autosampler rinsing solution: | Rinsing Solution (55007) |
Mass transitions of analytes and internal standards
Amino acids | Neutral Loss Scan 102 |
Alanine | 146 > 44 |
Alanine-D4 | 150 > 48 |
Aspartic acid | 246 > 144 |
Aspartic acid-D3 | 249 > 147 |
Glutamic acid | 260 > 158 |
Glutamic acid-D5 | 265 > 163 |
Leucine | 188 > 86 |
Leucine-D3 | 191 > 89 |
Methionine | 206 > 104 |
Methionine-D3 | 209 > 107 |
Phenylalanine | 222 > 120 |
Phenylalanine-D5 | 227 > 125 |
Tyrosine | 238 > 136 |
Tyrosine-D4 | 242 > 140 |
Valine | 174 > 72 |
Valine-D8 | 182 > 80 |
MRM | |
Arginine | 231>70 |
Arginine-D7 | 238 > 77 |
Citrulline | 232 > 113 |
Citrulline-D2 | 234 > 115 |
Glycine | 132 > 76 |
Glycine-13C2-15N | 135 > 79 |
Ornithine | 189 > 70 |
Ornithine-D6 | 195 > 76 |
Proline | 172 > 116 |
Proline-D7 | 179 > 123 |
SUAC | 211 > 109 |
SUAC-13C5 | 216 > 114 |
(Acyl-)carnitines | Parent Ion Scan 85 |
Carnitine | 218 > 85 |
Carnitine-D9 | 227 > 85 |
C2-Carnitine | 260 > 85 |
C2-Carnitine-D3 | 263 > 85 |
C3-Carnitine | 274 > 85 |
C3-Carnitine-D3 | 277 > 85 |
C4-Carnitine | 288 > 85 |
C4-Carnitine-D3 | 291 > 85 |
C5-Carnitine | 302 > 85 |
C5-Carnitine-D9 | 311 > 85 |
C5DC-Carnitine | 388 > 85 |
C5DC-Carnitine-D6 | 394 > 85 |
C6-Carnitine | 316 > 85 |
C6-Carnitine-D3 | 319 > 85 |
C8-Carnitine | 344 > 85 |
C8-Carnitine-D3 | 347 > 85 |
C10-Carnitine | 372 > 85 |
C10-Carnitine-D3 | 375 > 85 |
C12-Carnitine | 400 > 85 |
C12-Carnitine-D3 | 403 > 85 |
C14-Carnitine | 428 > 85 |
C14-Carnitine-D3 | 431 > 85 |
C16-Carnitine | 456 > 85 |
C16-Carnitine-D3 | 459 > 85 |
C18-Carnitine | 484 > 85 |
C18-Carnitine-D3 | 487 > 85 |
Order no. | Product |
---|---|
55000 | MassChrom® Reagent kit for the determination of Amino Acids and Acylcarnitines from dried blood spot samples (LC-MS/MS) with regular 96 well plates. For 960 analyses. |
55000/F | MassChrom® Reagent kit for the determination of Amino Acids and Acylcarnitines from dried blood spot samples (LC-MS/MS) with 96 well filter plates. For 960 analyses. |
Separately available components | |
55001 | Mobile Phase, 1000 ml |
55002 | Mobile Phase, 10 x 1000 ml |
55004 | Internal Standard (lyoph.), 4 x 50 ml |
55044 | Internal Standard for Succinylacetone, 3 x 50 ml |
55004/T | Internal Standard (lyoph.), 50 ml |
55044/T | Internal Standard for Succinylacetone, 50 ml |
55005 | Derivatisation Reagent, 30 ml |
55006 | Reconstitution Buffer, 100 ml |
55007 | Rinsing Solution, 1000 ml |
55008 | Extraction Buffer, 200 ml |
55010 | 96-Well-Plates, 10 pcs. |
55011 | Protective sheets for 96 well plates, 10 pcs. |
55057 | 96-Well-Filter-Plates, 5 pcs. |
Succinylacetone Upgrade Set | |
55111 | Succinylacetone Upgrade Set for 960 analyses, consisting of: |
Internal Standard for Succinylacetone, 3 x 50 ml | |
Reconstitution Buffer, 100 ml | |
Startup accessories | |
55013 | Pierceable adhesive seal for 96 well plates, 10 pcs. |
55015 | Restrictor Capillary, 1 pc. |
55016 | Adapter collar for centrifugation, 2 pcs. |
55098 | Tuning Mix for Succinylacetone, 1 ml |
55099 | Tuning Mix incl. all analytes, 2 ml |
55033 | PEEK prefilter, 5 µm, 5 pcs. |
15010 | PEEK prefilter housing, 1 pc. |
Chromsystems quality control material (filter spot): | |
0191 | MassCheck® Amino Acids, Acylcarnitines Dried Blood Spot Control, Bi-Level (I+II), 2 x 3 Spots |
0192 | MassCheck® Amino Acids, Acylcarnitines Dried Blood Spot Control, Level I, 1 x 3 spots |
0193 | MassCheck® Amino Acids, Acylcarnitines Dried Blood Spot Control, Level II, 1 x 3 spots |