♦ Air Removal Test Pack – Intended to evaluate the performance of the Air Removal (vacuum) system of Pre-Vacuum equipped Steam Sterilizers.
空气排出测试包-评估预真空蒸汽灭菌器的空气排除(真空)性能。
♦ Pre-assembled, disposable test pack. After exposure, unwrap pack and remove the indicator sheet for your records.
预装配的一次性测试包。灭菌后,解开包裹,取出试纸记录结果。
♦ 510(k) notification and meets AAMI, EN-867-3 and ISO 11140 air removal test.
符合510(k),AAMI,EN-867-3和ISO11140空气排除测试标准。
♦ Lead Free Ink on Indicator Sheet – Safer for medical personnel and the environment.
试纸采用无铅油墨-保证医疗人员和环境的安全。
♦ Distinct, non-reversible pink to black color change – easy to interpret.
*的不可逆的粉红色到黑色的变化-容易判读。
灭菌前 灭菌成功 灭菌失败
Test sterilizer once a day.
Test at the beginning of each day. A shortened cycle should be used to preheat the system (i.e. omit the drying phase). If the sterilizer is in continual use, then testing can be performed at any time, but it should be tested at the same time every day.
Place the pack on the bottom rack in the front section of the sterilizer, near the door and over the drain.
The test is most accurate when nothing else is in the chamber, so do not use this cycle to sterilize items or packs. Other items in this cycle may invalidate the test.
Run a pre-vacuum cycle following the parameters printed on the indicator instructions. You may eliminate the drying phase to save time without affecting the test.
After the cycle is complete, remove the pack and open to examine the indicator sheet. CAUTION! Test pack may still be hot so use caution when handling to avoid injury.
Record the necessary information on the sheet (or in a log book) and store for your records.
The filter paper pack can be discarded into a paper recycle bin.